Mahapatra, P.K. and Walia, M.S. and Pandey, Laxmi and Singh, Mewa and Jain, S.C. and Singla, M.L. (2004) VisiDAQ3.1 based optoelectronic system for the quantitative analysis of inorganic Phosphate-Phosphorous: absorption spectroscopy. Journal of Scientific & Industrial Research, 63 (6). pp. 509-511. ISSN 0022-4456
|
PDF
VisiDAQ3.1 based optoelectronic.pdf Download (135Kb) | Preview |
Abstract
The paper describes an optoelectronics system using ADAM-4011D data acquisition module and VisiDAQ 3.1 software for measurement of concentration, absorbance, and transmission of various compounds based on Lambert’s-Beer’s Law. VisiDAQ is data acquisition and control application software, which can be used in applications indulging automation and process control which reduces the hardware. ADAM-4011D is a data acquisition module, which converts analog output of I-V converter to digital output. This digital output is fed to the computer through serial port, which is processed by VisiDAQ software.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | ADAM-4011D module, visiDAQ 3.1 software;Photo detector |
Subjects: | CSIO > Chemistry |
Depositing User: | Ms. Narinder Kaur |
Date Deposited: | 04 Apr 2012 16:07 |
Last Modified: | 11 Apr 2012 12:13 |
URI: | http://csioir.csio.res.in/id/eprint/23 |
Actions (login required)