Kumar, Neelesh and Pankaj, Dinesh and Mahajan, Ankit and Kumar, Amod and Sohi, B.S. (2009) Evaluation of normal gait using electro-goniometer. Journal of Scientific & Industrial Research, 68 (8). pp. 696-698. ISSN ISSN: 0022-4456 (Print) ISSN: 0975-1084 (Online)
|
PDF
JSIR 68(8) 696-698.pdf Download (49Kb) | Preview |
Abstract
A potentiometer-based electro-goniometer has been developed to measure knee angle of normal healthy individuals (5),who were instructed to walk slow, normal and brisk for walking distance of 17 m. Data acquisition and analysis on acquired datawere done using LabVIEW. For individuals with different physical parameters, it was found that frequency (steps/min) remains the same irrespective of varying velocities. Step length reduces as individuals walk without shoes.
Item Type: | Article |
---|---|
Uncontrolled Keywords: | Gait cycle, Goniometer, Knee flexion angle, Step length, Stride length |
Subjects: | CSIO > Medical Instrumentation |
Divisions: | Medical Instrumentation |
Depositing User: | Ms T Kaur |
Date Deposited: | 22 Jan 2012 14:28 |
Last Modified: | 22 Jan 2012 14:28 |
URI: | http://csioir.csio.res.in/id/eprint/236 |