Design and implementation of controlled drug infusion system.

Gupta, R.C. and Taneja, S.R. and Thariyan, K.K. and Kumar, Sanjeev (2005) Design and implementation of controlled drug infusion system. Journal of Scientific and Industrial Research, 64 (10). pp. 761-766. ISSN 0022-4456

[img]
Preview
PDF
Design and implementation of controlled drug infusion system JSIR 64(10) 761-766.pdf

Download (238Kb) | Preview
Official URL: http://nopr.niscair.res.in/bitstream/123456789/528...

Abstract

An inexpensive controlled drug infusion system based around 80C85 microprocessor with RAM, EPROM and peripheral interface devices is developed. In this system, the drop rate selection is done by BCD thumb wheel switches from 1 to 99 drops/min, LCD display is used for displaying the parameters settings, number of drops infused and various alarms messages for patient safety and unique drop sensing assembly is incorporated for accurate detection of falling drops in drip chamber. The prototype has been tested successfully on neonates.

Item Type: Article
Uncontrolled Keywords: Drug infusion; Drop sensing assembly; Microprocessor; Alarms; I V set
Subjects: CSIO > Medical Instrumentation
Depositing User: Ms. J Shrivastav
Date Deposited: 03 Apr 2012 16:02
Last Modified: 03 Apr 2012 16:02
URI: http://csioir.csio.res.in/id/eprint/27

Actions (login required)

View Item View Item