Design and development of a unique drop sensing unit for infusion pump.

Thariyan, K.K. and Verma, S. and Taneja, S.R. and Gupta, R.C. and Ahluwalia, S.S. (2002) Design and development of a unique drop sensing unit for infusion pump. Journal of Scientific & Industrial Research, 61 (10). pp. 798-801. ISSN 0022-4456

Full text not available from this repository.
Official URL: http://www.niscair.res.in/sciencecommunication/res...

Abstract

In the present set-up the precise assembly of drop sensor unit has three IR-LED/detector pairs placed at different planes across drip chamber to detect the falling drop. The prismatic effects due to drops sticking to the walls in front of one pair of transmitter/detector is compensated by other two pairs. This increases a reliability and dependability of the overall system. The unit can accommodate variable sizes of drip chambers available in the market. Details of the above unit are given and the overall performance improvement due to the present device are also discussed.

Item Type: Article
Subjects: CSIO > Agri - Instrumentation
Depositing User: Ms. J Shrivastav
Date Deposited: 29 Mar 2012 12:55
Last Modified: 29 Mar 2012 12:55
URI: http://csioir.csio.res.in/id/eprint/162

Actions (login required)

View Item View Item